Meaning
High-resolution volumetric measurements capture the specific absorption rate distribution within a localized region of a tissue-equivalent phantom. The dasy zoom scan identifies the maximum energy deposition by sampling a defined cube of points around a detected hot spot. This automated process uses a robotic arm to position a miniature field probe with precision.
Each measurement point contributes to a three-dimensional map of the electric field intensity.
Spatial Sampling
Grid density determines the ability of the system to resolve rapid changes in energy levels near the surface of the device. A dasy zoom scan typically employs a five by five by seven or seven by seven by seven point matrix to cover the area of interest. These points are closer together than those used in the initial area scan.
Data Extrapolation
Mathematical algorithms calculate the field values at the surface of the phantom where the physical probe cannot reach. Because the probe sensor has a finite offset from its tip, the dasy zoom scan must estimate the peak values using polynomial regression or spline interpolation. This step ensures the final report accounts for the highest possible exposure level.
The calculation remains valid even when the gradient is steep.
Measurement Accuracy
Validating the system using a reference dipole verifies that the mechanical components and software are functioning within the allowed uncertainty budget. The dasy zoom scan provides the raw data needed for the averaging calculation of the energy absorbed by a specific mass of tissue.