Meaning
Signal power measurements for complex modulated waves require calculated averages that reflect the true heating value of the waveform. Receiver instruments use an RMS average detector to measure power by squaring the input voltage, averaging it, and calculating the square root. This process yields an accurate power value regardless of the modulation format.
The technique is less effective for single unmodulated carrier waves where simpler detectors suffice.
Signal Integration
Averaging times must exceed the period of the lowest modulation frequency to ensure a stable measurement. In an RMS average detector, this is achieved by adjusting the video bandwidth of the analyzer. Too short an averaging time causes the output to fluctuate.
This stabilizes the reading.
Power Measurement
Modern signals with high peak-to-average power ratios defeat traditional diode detectors. Incorporating an RMS average detector into the signal analyzer ensures that the reported power matches the physical energy delivered to the antenna. This accuracy is essential for compliance testing.
It prevents false failures.
Instrument Sweep
Sweeping across a frequency span requires the detector to dwell long enough at each point to compute the average. When using an RMS average detector, the sweep time of the analyzer increases to accommodate the calculation window. Adjusting this parameter balance prevents measurement errors due to under-sampling.
This ensures that transient peaks do not skew the average. It also provides repeatable traces during high-speed tests.