Meaning
Physical dimensions of the sensing elements determine the minimum distance at which two distinct points in an electromagnetic field can be distinguished. High probe spatial resolution is necessary to accurately map the steep field gradients found near the surface of modern high-frequency antennas. If the probe is too large, it will average the field over its own volume and underestimate the true peak intensity.
This parameter is a fundamental characteristic of the measurement hardware.
Sensor Dimension
Miniature dipoles arranged in a triangular or orthogonal configuration form the sensing element of the measurement probe. To achieve high probe spatial resolution, manufacturers reduce the length of these dipoles to less than three millimeters.
Gradient Detection
Sharp changes in energy density occur within a few millimeters of the transmitter. A system with superior probe spatial resolution captures these variations without the blurring effects of spatial averaging. This precision is necessary for devices operating in high frequency bands where the wavelength is short.
Field Mapping
Software corrections compensate for the displacement of the sensors from the geometric center of the probe housing. Even with these corrections, the probe spatial resolution sets the limit for how fine the measurement grid can be. A higher resolution reduces the uncertainty in the final calculated absorption rate.