Meaning
Probe accuracy measurements determine the geometric distortion introduced by a coordinate measuring machine sensor when tracing a reference sphere. The single stylus form error is the difference between the maximum and minimum radial deviations measured during a spherical scan using a single probe tip. This metric isolates the physical roundness of the stylus tip and the kinematic behavior of the probe head from other machine error sources.
It is a critical performance value that defines the limit of detail a coordinate measuring machine can accurately measure on a component surface.
Metrology Testing
Testing involves measuring a certified reference sphere of known diameter. The probe takes a set number of points across the spherical surface to establish the single stylus form error. This process is automated within the calibration software of the coordinate measuring machine.
Influence Factor
Stylus wear and surface contamination directly degrade the accuracy of the probe. The single stylus form error increases when the probe sphere has microscopic flat spots or holds micro-debris from repeated scanning of soft metals like aluminum. This error causes false variations in the measured form of manufactured parts.
Regular cleaning of the stylus tip with isopropyl alcohol minimizes this source of variation.
Quality Standard
Metrology laboratories trace this error to verify that the probe module is functioning within its design limits. The single stylus form error is recorded on the daily machine check sheet before any product measurements are performed. This record provides a continuous history of probe health and warns of impending module failure.
It is the primary indicator used to decide when a stylus must be replaced.