Meaning
Calibration standards in radiated power measurements use a known, highly characterized antenna to replace the device under test during testing. Utilizing a substitute radiator allows laboratories to determine the effective radiated power of a wireless device. This method replaces the unknown emissions with a controlled signal of known strength.
It is an accepted technique for regulatory compliance testing.
Measurement Procedure
Testing procedures in anechoic chambers require a two-step process to ensure accurate results. First, the transmitter under test is measured, and then it is replaced with the substitute radiator at the exact same physical position. A signal generator drives this reference antenna to replicate the received power level.
This procedure calibrates the path loss of the measurement chamber.
Reference Calibration
The reference antenna must have a fully documented radiation pattern and gain profile. This documentation ensures the accuracy of the substitute radiator measurements.
Error Reduction
Path loss and cable attenuation can vary with temperature and frequency, introducing errors into radio frequency measurements. By using a substitute radiator, these variables are cancelled out because both the test device and the reference antenna experience the same environmental conditions. This method is used for compliance testing of transmitters where direct cable connection is impossible.
The resulting measurements are highly reliable and can be reproduced across different testing laboratories, which is essential for global product certification. Accurate data protects manufacturers from the risk of regulatory non-compliance.