Meaning
Baseline sensitivity of a spectrum analyzer defined by its internal thermal noise floor. Most specifications define the displayed average noise level with the input terminated and the resolution bandwidth set to its minimum. This value determines the smallest signal the instrument can distinguish from the background.
Sensitivity Threshold
Measurement accuracy for low-level emissions depends on the distance between the signal peak and the noise floor. Adjusting the resolution bandwidth reduces the displayed average noise level by narrowing the noise integration window. A wider bandwidth captures more thermal energy and raises the floor.
Instrument Calibration
Attenuator settings at the input stage influence the visible noise on the screen. While the actual internal noise remains constant, the displayed average noise level appears to rise when input attenuation increases because the analyzer compensates for the lost signal. This shift can mask low-amplitude spurious emissions during electromagnetic interference testing.
Verification Protocol
Pre-amplifiers are often integrated into the signal path to lower the effective floor for sensitive measurements. Engineers verify the displayed average noise level before conducting compliance scans to ensure the equipment meets the dynamic range requirements of the test standard. If the floor is too high, narrow signals might remain hidden beneath the grass.
This verification provides a baseline for all subsequent amplitude readings.