Meaning
Field probe technology designed for high-precision near-field power density measurements evaluates radio frequency exposure from millimeter-wave transmitters. The EUmmWVx probe features two sensor dipoles optimized for frequencies from the low gigahertz range up to one hundred gigahertz. This sensor configuration minimizes field disturbance during close-proximity measurements of module antennas.
Sensor Technology
Advanced sensor integration employs micro-machined elements to achieve minimal scattering and high sensitivity. Each dipole in the EUmmWVx assembly is coupled to a zero-bias Schottky diode that generates a voltage proportional to the local field. This design eliminates the need for active coaxial cabling, reducing the risk of stray RF currents corrupting the measurement.
Measurement Protocol
Automated scan sequences reposition the active sensor across a predetermined grid above the integrated device. To determine compliant exposure levels, the software utilizes EUmmWVx readings to reconstruct the three-dimensional power density distribution. The test sequence records phase and amplitude at varying heights, allowing the algorithm to calculate both radiative and reactive near-field components.
Calibration Procedure
Traceable reference sources establish the correction factors required for accurate absolute measurements. Because slight changes in probe geometry alter the output, each EUmmWVx undergoes periodic validation in a specialized waveguide chamber. The calibration process matches the sensor voltage to a known power level across all target bands.
This validation guarantees that the field readings taken during system compliance testing are reproducible and accurate to within defined limits. Testing laboratories record these correction factors in the official compliance folder before the product goes to market.