Meaning
Active gain blocks placed before a measurement receiver improve system sensitivity by overcoming cable and instrumentation noise. In radiated emissions setups, low noise preamplifiers are inserted between the receive antenna and the analyzer to boost weak signal levels. This amplification helps ensure that low level emissions are not lost in the high noise floor of the spectrum analyzer.
Signal Amplification
High gain combined with a very low internal noise figure is the primary requirement for these amplifiers. When low noise preamplifiers are added to the signal path, they must have a flat frequency response across the test range. This flatness simplifies the calibration and correction process.
Noise Figure
The overall noise figure of the test system is dominated by the noise figure of the first amplifier in the chain. Minimizing this first stage noise is essential for detecting low amplitude spurious emissions.
Practical Integration
Designers must balance the preamplifier gain against the risk of receiver overload. If the amplifier saturates, it will generate intermodulation products that appear as false emission peaks on the analyzer. Bandpass filters can be placed before the amplifier to protect the input stage from strong out-of-band signals.
This combination is common in microwave measurements where high path losses must be offset.