Meaning
A linearity test quantifies the third-order intermodulation products generated when two closely spaced frequencies pass through a nonlinear device. Performing a dual tone im3 measurement involves injecting two fundamental signals of equal power into a component and observing the resulting distortion peaks at specific offsets. These peaks appear at 2f1 minus f2 and 2f2 minus f1, where f1 and f2 are the input frequencies.
The resulting value defines the upper limit of the dynamic range for receivers and power amplifiers. High linearity ensures that signals remain distinguishable in crowded spectrum environments.
Distortion Metric
The ratio between the power of the fundamental signals and the power of the intermodulation products defines the linearity of the system. Dual tone im3 measurement provides the data necessary to calculate the third-order intercept point.
Frequency Selection
Choosing the appropriate spacing between the two test tones reveals the bandwidth limitations of the nonlinear behavior. A dual tone im3 measurement uses spacing that matches the actual channel widths of the target communication standard. Wide spacing identifies memory effects in the power amplifier, while narrow spacing focuses on static nonlinearities.
Engineers select these frequencies to avoid overlap with existing harmonics or system noise floors.
Power Threshold
Testing must occur at power levels where the device is operating within its normal range to avoid saturation effects. The dual tone im3 measurement yields valid results only when the output power of the distortion products follows a predictable three-to-one slope relative to the input power. Attenuators are placed before the spectrum analyzer to prevent the test equipment itself from introducing additional intermodulation.
This step ensures the measured distortion originates solely from the device under test.