Meaning
Calibration sequences establish the operational slope relating raw sensor electrical output to physical units of measurement across full scale. Executing span calibration requires applying a certified reference mass near maximum capacity to adjust system gain parameters. Regular adjustment eliminates gain drift caused by component aging and temperature shifts.
The procedure applies specifically to linear sensor response regions after zero point offset adjustment has been completed.
Gain Adjustment
Digital signal processors recalculate scaling factors when a known reference weight rests on the measurement pan. Internal motor-driven calibration weights execute automated span calibration routines inside sealed industrial modules. Test station software verifies that the calculated scale factor falls within factory specification limits before certifying the station.
Continuous monitoring flags gain shifts that exceed pre-set process control limits.
Reference Standard
Certified calibration weights with traceable mass values provide the physical reference needed for slope verification. Class E2 or F1 weights satisfy strict tolerance requirements for high-resolution production balances.
Environmental Sensitivity
Temperature variations modify sensor sensitivity by altering strain gauge gauge factors or magnetic flux density in restoring coils. Thermal equilibration prevents errors during high-accuracy reference weight applications. Automated test executive scripts log ambient temperature along with gain adjustments to track long-term line stability.
Drift limits trigger re-calibration protocols when environmental conditions cross defined thresholds.